Worked example: an MnPt film, fitted with RBS and checked with PIXE

This walk-through fits a real measurement with RBS alone, then holds the result against the PIXE spectrum measured in the same run. Unlike Fe and Ni, Mn and Pt are far apart in RBS, so RBS gives the composition by itself. PIXE is not fitted at all: it only checks that every element is where the RBS sample puts it, in about the right amount. It uses the two files in examples/:

  • MnPt.RBS, the RBS spectrum: an RC43 acquisition macro, 1.9 MeV ⁴He⁺, 10 µC, detector at 170°, sample tilted to THETA -9;
  • MnPt.PIX, the PIXE spectrum measured in the same run, with the same charge.

The sample is Si / SiO₂ / Ru / MnPt / Ru: an MnPt film between two thin Ru layers, on the oxide of a Si wafer. The question is the MnPt composition and the layer thicknesses.

Every command below is typed at the prompt it is shown with. The outputs are pyRUMP's own, from this data.

1. Load both spectra

PAIR ON makes reading x.RBS also read x.PIX from the same folder into the same buffer, so the two spectra share the beam, the tilt and the charge:

Your wish? cd examples
Your wish? pixe pair on
  pair on
  ! THETA 0 from the RBS buffer (PAIR ON): beam 0 deg, X-rays 45 deg to the sample normal
Your wish? xeq MnPt.RBS
buffer 1 emptied
  c:\RBS\data\2026\08\MA8408.RBS
  MA8408.RBS  170 Degree RBS LT =  951.165 RT  953.288 Gain  2
  16:43:33  08-19-2026
  1.59 keV/channel, offset 48 keV
  MeV = 1.901
  beam Z=2 mass=4.0026 charge state 1
  geometry = general
  theta = -9 deg
  phi = 11 deg
  psi = 20 deg
  omega = 2.7 msr
  first = 0
  FWHM = 15 keV
  current = 10.94 nA
  charge = 10 uC
2048 points entered into buffer 1
  PAIR: MnPt.PIX -> PIXE of this buffer

The sample's name, MA8408, is the first word of the spectrum's identifier. SNAP names its files after it (step 6).

2. Read the RBS and PIXE spectra

PIXE turns PIXE on, so PLOT shows the PIXE spectrum in its own window next to the RBS one. Zoom both on what matters, and ask ELEMENT where each element should show up:

Your wish? pixe
PIXE enabled: PLOT and COMPARE now show PIXE too (DISABLE turns it off)
PIXE Command: region 100 1200
  region 100 to 1200
PIXE Command: return
Your wish? plot 1
Your wish? region 300 1130
Your wish? element Pt Ru Mn Si O
  Pt  Z=78  Mass=195.090  K(ion)=0.9219  Energy=  1752.5 keV  Channel=1072.026
      X-rays  Lα 9.435 keV ch 938.2   Lβ1 11.071 keV ch 1100.3   Lβ2 11.248 keV ch 1117.9
              Lγ1 12.942 keV ch 1285.6   Mα 2.050 keV ch 206.8   Mβ 2.127 keV ch 214.5
  Ru  Z=44  Mass=101.070  K(ion)=0.8547  Energy=  1624.7 keV  Channel= 991.656
      X-rays  Kα 19.235 keV ch 1908.8   Lα 2.558 keV ch 257.2   Lβ1 2.683 keV ch 269.6
              Lβ2 2.836 keV ch 284.7   Lγ1 2.965 keV ch 297.5
  Mn  Z=25  Mass= 54.938  K(ion)=0.7488  Energy=  1423.5 keV  Channel= 865.090
      X-rays  Kα 5.895 keV ch 587.7   Kβ 6.490 keV ch 646.6   Lα 0.637 keV ch 66.9
              Lβ1 0.648 keV ch 68.0
  Si  Z=14  Mass= 28.086  K(ion)=0.5663  Energy=  1076.4 keV  Channel= 646.819
      X-rays  Kα 1.740 keV ch 176.1   Kβ 1.836 keV ch 185.7
  O   Z= 8  Mass= 15.999  K(ion)=0.3631  Energy=   690.2 keV  Channel= 403.874
      X-rays  Kα 0.525 keV ch 55.8

For each element ELEMENT gives the RBS surface edge and, with PIXE on, the main X-ray lines, and marks both on the plots.

The MnPt RBS spectrum as loaded, with ELEMENT's surface edges for Pt, Ru, Mn, Si and O ticked along the bottom

The RBS spectrum, from high to low energy:

  • a Pt peak near channel 1055, a little below the Pt surface edge: the MnPt film lies under the top Ru layer;
  • two Ru peaks, near 985 and 958. The higher one, at the Ru surface edge, is the Ru layer on top; the lower one is the Ru layer under the MnPt, its ions having lost energy crossing the film on the way in and out;
  • an Mn peak near 845, below the Mn surface edge for the same reason as the Pt;
  • the Si step near 600 and the O peak near 365, both well below their surface edges: the oxide and the wafer lie under the metal.

Mn and Pt are some 200 channels apart, so RBS separates them easily: their peak areas give the composition, and their widths the film thickness. That is the difference from the FeNi example, where Fe and Ni overlap and the composition has to come from PIXE.

The MnPt PIXE spectrum as loaded, with ELEMENT's X-ray lines marked: Si K, Pt M, Ru L, Mn Kα and Kβ, Pt L

The PIXE spectrum has a line for every element of the sample except O, whose K line at 0.5 keV the detector's window and filter stop: Si K from the wafer and the oxide, Pt M and Ru L just above it, Mn Kα and Kβ, and Pt Lα and Lβ. There is no line that the sample doesn't explain, so no element is missing from it.

Unlike the FeNi data, this spectrum needs no PIXE setup: pyRUMP's built-in PIXE defaults are this detector's, measured on this very spectrum, so the markers sit on the peaks as they are.

3. Describe the sample in SIM, layer by layer

Layer 1 is the surface, the layer the beam meets first. Each layer gets a thickness and a composition; the composition list ends with /. The unit follows the number: A (Å, the default), nm, um, or a compound from density.tab.

Your wish? sim
SIM Command: layer 1
  you are now working on a fresh layer # 1 (of 0)
SIM Command: thickness 40 A
SIM Command: composition Ru 1 /
SIM Command: next
  you are now working on a fresh layer # 2 (of 1)
SIM Command: thickness 300 A
SIM Command: composition Mn 3 Pt 1 /
SIM Command: next
  you are now working on a fresh layer # 3 (of 2)
SIM Command: thickness 40 A
SIM Command: composition Ru 1 /
SIM Command: next
  you are now working on a fresh layer # 4 (of 3)
SIM Command: thickness 300 SIO2
SIM Command: composition Si 1 O 2 /
SIM Command: next
  you are now working on a fresh layer # 5 (of 4)
SIM Command: thickness 5 um
SIM Command: composition Si 1 /
SIM Command: show
    1            40 A        Ru 1        [29 /CM2  Ru 29.03]
    2           300 A        Mn 3 Pt 1   [231 /CM2  Mn 173.35 Pt 57.78]
    3            40 A        Ru 1        [29 /CM2  Ru 29.03]
    4           300 SIO2     Si 1 O 2    [198 /CM2  Si 66 O 132]
 >  5             5 um       Si 1        [24888 /CM2  Si 24888.43]
  maxpth 1000   straggle 0   multiple 0   absorber 0
SIM Command: save MnPt
wrote MnPt.lcm
SIM Command: return
  • Layers 1–3, Ru / MnPt / Ru: rough starting values for the fit. Mn 3 Pt 1 is a starting ratio; in MODE COMP the numbers are relative amounts.
  • Layer 4, SiO₂, 300 Å, also fitted. The unit is SIO2, not A: Å are converted to atoms/cm² with an atomic density, and for a compound A takes it from the pure elements (300 Å of SiO₂ would be only 135 × 10¹⁵ at/cm² instead of 198). SIO2 is Å at silica's own density, from density.tab. For Ru and MnPt the elements' densities are right.
  • Layer 5, the Si wafer: the last layer is the substrate. It only has to be thicker than the beam can see; 5 µm is plenty for 1.9 MeV He, and it also gives PIXE the full Si K yield of the wafer.

SIM SAVE MnPt writes the description to MnPt.lcm — the file of that name in examples/ is exactly this — and SIM GET MnPt reads it back.

4. Fit RBS in PERT

Your wish? pert
PERT Command: window 400 1200
  error windows [1] 400-1200
PERT Command: fwhm
  varying fwhm
PERT Command: correction
  varying correction
PERT Command: kev(0)
  varying kev(0)
PERT Command: thickness 1
  varying layer 1 thickness
PERT Command: thickness 2
  varying layer 2 thickness
PERT Command: thickness 3
  varying layer 3 thickness
PERT Command: composition 2 Mn
  varying layer 2 composition Mn
PERT Command: kev/ch
  varying kev/ch
PERT Command: thickness 4
  varying layer 4 thickness
PERT Command: parms
  mode        multiple variable
  autocmp     off
  autosnap    off
  highlight   on
  error win   [1] 400-1200
  norm win    (none)
  PIXE win    (none -- RBS only)
  varying:
    [1] fwhm
    [2] correction
    [3] kev(0)
    [4] layer 1 thickness
    [5] layer 2 thickness
    [6] layer 3 thickness
    [7] layer 2 composition Mn
    [8] kev/ch
    [9] layer 4 thickness

What each setting does:

  • WINDOW 400 1200: the RBS channels fitted: the Si step, and the Mn, Ru and Pt peaks. The O peak, below channel 400, is left out; the SiO₂ shows in the window through where the Si step sits.
  • FWHM, KEV(0), KEV/CH: the RBS resolution and energy calibration. The values in the file header are only a starting point.
  • CORRECTION: the dose. The height of the Si plateau depends only on the dose and on silicon's stopping, so the window's Si plateau pins it. CORRECTION applies to the PIXE spectrum too, which step 5 relies on.
  • THICKNESS 1, 2, 3, 4: the top Ru, the MnPt, the bottom Ru and the SiO₂. The substrate stays as it is.
  • COMPOSITION 2 Mn: Mn's content against Pt, which stays at 1 as the reference.

PARMS shows PIXE win (none -- RBS only): with no PIXWIN, nothing is fitted to PIXE. GO fits:

PERT Command: go
  Fitting MA8408: Si [5um] - SiO2 [300SIO2] - Ru [40A] - Mn3Pt [300A] - Ru [40A]

  fit took 0.43 s

  reduced chi-square 4.2900 on 792 dof   (was 141.5915)
  66 evaluations, converged
  fwhm                              19.2299  +/- 0.1166   (was 15)
  correction                       0.855636  +/- 0.001678   (was 1)
  kev(0)                            47.3853  +/- 0.1454   (was 48)
  layer 1 thickness                    40 A  +/- 0.3326 A   (was 40 A)   [29 /CM2]
  layer 2 thickness                   314 A  +/- 0.7404 A   (was 300 A)   [242 /CM2]
  layer 3 thickness                    41 A  +/- 0.3266 A   (was 40 A)   [30 /CM2]
  layer 2 composition Mn            2.74154  +/- 0.01577   (was 3)
  kev/ch                            1.59761  +/- 0.0001069   (was 1.59)
  layer 4 thickness                325 SIO2  +/- 6.634 SIO2   (was 300 SIO2)   [215 /CM2]

  MA8408: Si [5um] - SiO2 [325SIO2] - Ru [41A] - Mn2.74Pt [314A] - Ru [40A]
PERT Command: go
  ...
  reduced chi-square 3.9872 on 792 dof   (was 4.2731)
  ...
PERT Command: go
  ...
  reduced chi-square 3.7188 on 792 dof   (was 3.9872)
  ...
PERT Command: go
  Fitting MA8408: Si [5um] - SiO2 [325SIO2] - Ru [41A] - Mn2.75Pt [314A] - Ru [39A]

  fit took 0.27 s

  reduced chi-square 3.7029 on 792 dof   (was 3.7190)
  39 evaluations, converged
  fwhm                              20.3534  +/- 0.1265   (was 20.21)
  correction                       0.857688  +/- 0.001695   (was 0.857432)
  kev(0)                            45.4401  +/- 0.1634   (was 45.8488)
  layer 1 thickness                    39 A  +/- 0.3419 A   (was 39 A)   [29 /CM2]
  layer 2 thickness                   314 A  +/- 0.8059 A   (was 314 A)   [241 /CM2]
  layer 3 thickness                    41 A  +/- 0.3295 A   (was 41 A)   [30 /CM2]
  layer 2 composition Mn            2.74087  +/- 0.01609   (was 2.75193)
  kev/ch                            1.59945  +/- 0.000121   (was 1.59917)
  layer 4 thickness                325 SIO2  +/- 7 SIO2   (was 325 SIO2)   [215 /CM2]

  MA8408: Si [5um] - SiO2 [325SIO2] - Ru [41A] - Mn2.74Pt [314A] - Ru [39A]

Run GO until nothing moves, here four times. The composition and the thicknesses are settled after the first; the resolution and the calibration offset, which trade off against each other on the peak edges, take a few rounds more. A fifth GO changes nothing.

PERT Command: highlight off
  highlight off
PERT Command: return
Your wish? compare
Your wish? element Pt Ru Mn Si O

HIGHLIGHT OFF leaves the fit window unshaded, for a cleaner figure.

The MnPt RBS spectrum after the fit, with ELEMENT's surface edges marked

The fit follows the whole window, and below it the O peak too, which it was not fitted to. What remains is on the peak flanks: the data are above the simulation in the valley between the Ru and Pt peaks (around channel 1020) and on the low-energy side of the Mn peak (around 820). Some intermixing or roughness at the interfaces, which the model's sharp interfaces leave out, would do that.

Your wish? sim show
 >  1            39 A        Ru 1           [29 /CM2  Ru 28.54]
    2           314 A        Mn 2.74 Pt 1   [241 /CM2  Mn 176.83 Pt 64.52]
    3            41 A        Ru 1           [30 /CM2  Ru 29.71]
    4           325 SIO2     Si 1 O 2       [215 /CM2  Si 71.56 O 143.13]
    5             5 um       Si 1           [24888 /CM2  Si 24888.43]
  maxpth 1000   straggle 0   multiple 0   absorber 0
Layer Thickness Areal density (10¹⁵ at/cm²) Composition
Ru 39 Å 29
MnPt 314 Å 241 (Mn 177, Pt 65) Mn/Pt = 2.74 ± 0.02, i.e. Mn₇₃Pt₂₇
Ru 41 Å 30
SiO₂ 325 Å 215

The uncertainties are statistical only, as PERT reports them; the thicknesses in Å assume the elements' densities (silica's for the SiO₂). The fitted CORRECTION, 0.858, means about 17 % more ions reached the sample than the 10 µC the integrator recorded.

5. Check the result against PIXE

PIXE was turned on in step 2, so the COMPARE above also put the PIXE spectrum against the PIXE simulation of the fitted sample, at the dose the fit found. Nothing in it was fitted to PIXE:

The MnPt PIXE spectrum against the PIXE simulation of the RBS-fitted sample, with the X-ray lines marked

Every line is where the sample puts it, and on the log scale every line is about the right size. The residuals between the lines are the continuum background, and the peak tails, which pyRUMP doesn't model yet (the PIXE manual lists what is and isn't simulated): the data are above the simulation there.

For the sizes, LINES at the PIXE prompt lists the simulated counts of each line, largest first, with totals per shell:

Your wish? pixe
PIXE Command: lines
  element  line    E (keV)   sigma (b)   efficiency      counts
  Si       KL3      1.7400       307.1       0.0033   3.166e+04
  Si       KL2      1.7394       154.5       0.0033   1.584e+04
  Ru       L3M5     2.5585       140.7       0.1551        5359
  Pt       M5N7     2.0500       532.5       0.0286        4194
  Pt       M4N6     2.1270       250.8       0.0410        2834
  Ru       L2M4     2.6833        54.8       0.1978        2657
  Mn       KL3      5.8987       3.959       0.8582        2596
  ...
  totals:  Si K 5.057e+04, Ru L 9810, Pt M 9099, Mn K 4465, Pt L 577, Ru K 3.187, Pt K 0.0004016, Mn L 1.554e-22, O K 5.882e-28
PIXE Command: exportcmp mnpt-pixe.csv
wrote mnpt-pixe_pixe.csv: 2048 PIXE channels, reduced chi-square 16.2318 (1101 dof)

EXPORTCMP writes the measured and simulated counts channel by channel. Summed over each group of lines:

Lines Energy (keV) Measured Simulated Simulated / measured
Si K 1.55–1.95 49 263 50 613 1.03
Pt M 1.95–2.22 8 106 7 331 0.90
Ru L 2.45–3.10 9 977 10 694 1.07
Mn Kα 5.65–6.15 2 848 3 904 1.37
Mn Kβ 6.30–6.70 412 545 1.32
Pt Lα 9.20–9.70 347 361 1.04
Pt Lβ 10.80–11.50 181 165 0.91
  • Pt and Ru agree. The Pt L lines, the Pt M lines and the Ru L lines come out within 10 % of the measurement. That checks the Pt and Ru amounts RBS found, and the dose: the PIXE simulation runs at the CORRECTION the RBS fit set. Mn K is about 35 % too strong. RBS gives Mn as reliably as Pt here, so the excess is on the PIXE side: the setup's constants for K lines. The FeNi example, on the same detector, found H_K = 0.78 — K lines 1.3 times too strong — which is the same thing. With that H_K, Mn K agrees to within 7 %:
PIXE Command: h k 0.78
  h 0.78 1 1  ! K, L, M
Lines Simulated / measured, H_K = 1 H_K = 0.78
Mn Kα 1.37 1.07
Mn Kβ 1.32 1.03
Si K 1.03 0.80

Si K then drops 20 % low. The built-in filter thickness was measured on this spectrum's Si K with H_K = 1, so it already absorbs part of the same error. One H_K can't put both Si K and Mn K right until the filter and H are calibrated together on a standard.

So PIXE confirms the sample: every line is accounted for, Pt and Ru agree, and Mn agrees once the detector's K-line constant is set from another measurement. To get amounts from PIXE alone, calibrate H on a standard first (see Yield). Set H back before going on:

PIXE Command: h k 1
  h 1 1 1  ! K, L, M
PIXE Command: return

6. Keep the result: SNAP

SNAP (SNAPSHOT) writes the sample, the fit setup, both plots and a restore macro, named after the sample:

Your wish? snap

  updated MA8408.report
  wrote MA8408.pert
  wrote MA8408.lcm
  wrote MA8408_rbs.png
  wrote MA8408_pixe.png
  wrote MA8408_fit.xeq

MA8408.report gets the last fit and the sample, appended after any earlier ones; MA8408.lcm is the fitted sample. MA8408.pert holds everything step 4 typed — it is examples/MnPt.pert — and replays it in one line:

window 400 1200
fwhm
correction
kev(0)
thickness 1
thickness 2
thickness 3
composition 2 Mn
kev/ch
thickness 4
Your wish? sim get MnPt
Your wish? pert get MnPt go

PERT GET ... GO is one GO; type GO again until nothing moves, as in step 4. AUTOSNAP in PERT saves after every GO by itself, and pyrump MA8408_fit.xeq restores the whole session later; the FeNi example shows both.

See SIM, PERT, PIXE and SNAPSHOT in the manual for every option used here.